Discussion:
mesure avec nanoVNA
(too old to reply)
bp
2022-12-25 10:30:11 UTC
Permalink
Hello,

I bought this box:

https://www.aliexpress.com/item/4001095294804.html

I have no problem following the calibration procedure for a small
antenna with an sma connector

If I want to calibrate the antenna I have on the pylon (2m) and which
has a PL259 connector I have to ADD an SMA PL 259 adapter


During a last balancing manipulation I have to join by the cable given
in the box S11 and S21 (s0 and s1)

There are two bad solutions to calibrate the whole line:

a) remove the PL259 from the end of the cable to be able to connect the
SMA cable to the SMA input of the nanoVNA
it will miss the adapter capacitance

b) add to the PL259 a PL259 SMA adapter in order to be able to do the
last calibration manipulation but in this case I would have to remove
this adapter to fix the cable to the pl 259 of the antenna base


For the last balancing manipulation (connect So to S1 )

NanoVNA S0 ------> 20cm cable ------> SMA PL259 adapter -----> (this in
order to take into account during the measurement all the capacities of
the line) ---- ----> Pl259 SMA reducer ------> input S1 of the nanoVNA

When measuring the antenna, the last reducer (PL259 SMA) is no longer
used since I go directly into the PL259 of the antenna.


The calibration of this fact is no longer the right one because I have
to remove the last reducer (pl259 SMA )


What to do ?
Thank you

F11MYG

Use simple sentences because I translate with google




Bonjour,

j'ai acquis ce boitier :

https://www.aliexpress.com/item/4001095294804.html

Je n'ai pas de problème pour suivre la procédure de calibrage pour
une petite antenne avec un connecteur sma

Si je veux calibrer l'antenne que j'ai sur le mat et qui a un
connecteur PL259 je suis obligé d' AJOUTER un adaptateur sma PL 259


Lors d'une dernière manip d'équilibrage on doit joindre par le cable
donné dans la boite S11 et S21 (s0 et s1)

Il y a deux mauvaises solutions pour calibrer toute la ligne :

a) retirer la PL259 du bout du cable pour pouvoir faire la liaison SMA
cable à l'entrée SMA du nanoVNA

b)ajouter à la PL259 un adaptateur PL259 SMA afin de pouvoir faire la
dernièe manip de calibrage mais dans ce cas là je devrais retirer cet
adaptateur pour fixer le cable à la pl 259 de l'embase de l'antenne


Pour la dernière manip d'équilibrage (relier So à S1 )

NanoVNA S0 ------> cable de 20cm ------> adaptateur SMA PL259 ----->
(ceci afin de prendre en compte lors de la mesure toutes les capacités
de la ligne) --------> reducteur Pl259 SMA ------> entrée S1 du nanoVNA

Lors de la mesure de l'antenne le dernier reducteur (pl259 SMA ) n'est
plus utilisé puisque je rentre directement dans la PL259 de l'antenne.


Le calibrage de ce fait n'est plus le bon car je suis obligé de
retirer le dernier reducteur (pl259 SMA )


Que faire ?
Merci

F11MYG

Utilisez des phrases simples car je traduis avec google
Bernard
2022-12-27 16:51:22 UTC
Permalink
Bonjour,
Il me semble que le plus simple (c'est en tout cas ce que je fais) est
de calibrer l'instrument de SMA à SMA avec le jeu de charges fourni et
de considérer que les caractéristiques de l'antenne (impédance et
fréquence de résonance) ne sont que marginalement affectées par le
raccord PL/SMA, surtout en 2m. A cette fréquence, la longueur électrique
de la PL est très faible et je ne vois aucune raison pour considérer que
une PL a plus ou moins d'effet que 2cm de coax.
Sinon il faut construire un jeu de charges (50, open et CC) en PL pour
calibrer l'instrument au niveau du point de mesure de l'antenne.

Cordialement
Bernard (F1EVY)
Post by bp
Hello,
https://www.aliexpress.com/item/4001095294804.html
I have no problem following the calibration procedure for a small
antenna with an sma connector
If I want to calibrate the antenna I have on the pylon (2m) and which
has a PL259 connector I have to ADD an SMA PL 259 adapter
During a last balancing manipulation I have to join by the cable given
in the box S11 and S21 (s0 and s1)
a) remove the PL259 from the end of the cable to be able to connect the
SMA cable to the SMA input of the nanoVNA
it will miss the adapter capacitance
b) add to the PL259 a PL259 SMA adapter in order to be able to do the
last calibration manipulation but in this case I would have to remove
this adapter to fix the cable to the pl 259 of the antenna base
For the last balancing manipulation (connect So to S1 )
NanoVNA S0 ------> 20cm cable ------> SMA PL259 adapter -----> (this in
order to take into account during the measurement all the capacities of
the line) ---- ----> Pl259 SMA reducer ------> input S1 of the nanoVNA
When measuring the antenna, the last reducer (PL259 SMA) is no longer
used since I go directly into the PL259 of the antenna.
The calibration of this fact is no longer the right one because I have
to remove the last reducer (pl259 SMA )
What to do ?
Thank you
F11MYG
Use simple sentences because I translate with google
Bonjour,
https://www.aliexpress.com/item/4001095294804.html
Je n'ai pas de problème pour suivre la procédure  de calibrage  pour une
petite antenne avec un connecteur sma
Si je veux calibrer l'antenne que j'ai sur le mat et qui a un connecteur
PL259 je suis obligé d' AJOUTER un adaptateur sma PL 259
Lors d'une dernière manip d'équilibrage on doit joindre par le cable
donné dans la boite S11 et S21 (s0 et s1)
a) retirer la PL259 du bout du cable pour pouvoir faire la liaison SMA
cable à l'entrée SMA du nanoVNA
b)ajouter à la PL259 un adaptateur PL259 SMA  afin de pouvoir faire la
dernièe manip de calibrage mais dans ce cas là je devrais retirer cet
adaptateur  pour fixer le cable à la pl 259 de l'embase de l'antenne
Pour la dernière manip d'équilibrage (relier So à S1 )
NanoVNA S0  ------> cable de 20cm ------> adaptateur  SMA PL259 ----->
(ceci afin de prendre en compte  lors de la mesure toutes les capacités
de la ligne) --------> reducteur Pl259 SMA ------> entrée S1 du nanoVNA
Lors de la mesure de l'antenne le dernier reducteur  (pl259 SMA ) n'est
plus utilisé puisque je rentre directement dans la PL259 de l'antenne.
Le calibrage de ce fait  n'est plus le bon car je suis obligé de retirer
le dernier reducteur  (pl259 SMA )
Que  faire ?
Merci
F11MYG
Utilisez des phrases simples car je traduis avec google
bp
2022-12-27 22:09:41 UTC
Permalink
Post by bp
Bonjour,
Il me semble que le plus simple (c'est en tout cas ce que je fais) est de
calibrer l'instrument de SMA à SMA avec le jeu de charges fourni et de
considérer que les caractéristiques de l'antenne (impédance et fréquence de
résonance) ne sont que marginalement affectées par le raccord PL/SMA, surtout
en 2m. A cette fréquence, la longueur électrique de la PL est très faible et
je ne vois aucune raison pour considérer que une PL a plus ou moins d'effet
que 2cm de coax.
Sinon il faut construire un jeu de charges (50, open et CC) en PL pour
calibrer l'instrument au niveau du point de mesure de l'antenne.
Cordialement
Bernard (F1EVY)
Bonjour

Je pense comme vous que, effectivement cela n'entrainera pas une grosse
difference dans les résultats de la mesure et que cela ne vaut peut
être pas la peine de monter une usine à gaz .

Cordialement et avec mes remerciements

Bernard
F11MYG